TEMPSENS · TMC16-1
Schaffer Diaphragm Pressure Gauge
Temperature range: 100°C max.
Price on request
Details

TEMPSENS · TMC18-3
Temperature range: 10…150°C max.
Category: Industrial sensors · Gauges with diaphragm seals
| Characteristic | Value |
|---|---|
| Généralités | General |
| Conception | EN 837-3 |
| Tolérance | ±1.6% of full scale |
| Température ambiante | -40…+65°C [without dampening liquid] -20…+65°C [with dampening liquid] |
| Température de service | 10…150°C max. |
| Limites de pression | Steady pressure up to FS value Fluctuating pressure up to 90% of FS value Short time 1.3 x FS value for range up to 100 bar Short time 1.15 x FS value for range above 100 bar |
| Soudures | TIG argon arc welding |
| Version standard | Standard Version |
| Diamètre du cadran | DN100 |
| Étendue | -1…600 bar |
| Montage | Direct, Bottom connection |
| Raccord process | Without mounting flange |
| Seal filling fluid | Glycerine [10°C to 150°C] |
| Indice de protection | IP 65 |
| Exécution | Dry / Dry but fillable |
| Matériau | Material of construction |
| Élément sensible | Diaphragm |
| Matériau du boîtier et de la lunette | AISI 304 SS (Bayonet type) |
| Diaphragm, Bourdon tube & Shank | AISI 316L SS |
| Mécanisme | AISI 304 SS |
| Cadran | Aluminum, black graduation on white background |
| Aiguille | Fixed, aluminum, black powder coated |
| Joints | Neoprene |
| Hublot | Plexi glass |
| Versions remplies | Filled version |
| Glycérine | Glycerin |
| Hublot | Plexi glass |
| Liquide amortisseur | Glycerin 99.7% |
| Effet de la température | When temperature of the measuring system deviates from reference temperature (+20°C) : max error 0.4% / 10K of true scale value. |

TEMPSENS · TMC16-1
Temperature range: 100°C max.

TEMPSENS · TMC16-2
Temperature range: 100°C max.

TEMPSENS · TMC18-1
Temperature range: 10…150°C max.

TEMPSENS · TMC18-2
Temperature range: 10…150°C max.